RTD Thermometer / Dataloggers (PTD-12A)
Description
Specification: | |
Measurement Range: Thermocouple |
K Type:-200°C~1372°C (-328°F~2501°F) J Type:-210°C~1200°C (-346°F~2192°F) T Type:-250°C~400 °C (-418°F~752°F) E Type:-210°C~1000°C (-346°F~1832°F) R/S Type:0°C~1767°C (32°F~3212°F) N Type:-150°C~1300°C (-238°F~2372°F) |
Accuracy: (18°C ~ 28°C ambient) | K/J/E/T/N Type:±【0.05% reading + 0.3°C (0.6°F)】 【below-100°C (-148°F):K/J/T/E add 0.15% of reading ; N add 0.45% of reading】 R/S Type:±【0.05% reading + 1°C (1.8°F)】 |
Resolution: |
0.1 for K/J/T/E/N < +2000°C/°F ; K/J/T/E/N > -2000°C/°F ; R/S < +1000°C (1832°F) ; R/S > -1000°C (-1832°F); 1 for K/J/T/E/N > +2000°C/°F ; K/J/T/E/N < -2000°C/°F ; R/S > +1000°C (1832°F) ; R/S < -1000°C (-1832°F) |
Temperature Coefficient | 0.01% of reading + 0.03°C per °C (0.06°F per °F) outside the specified +18°C to 28°C (+64°F to +82°F) range 【below-100°C (-148°F):K/J/T/E add 0.04% of reading;N add 0.08% of reading】 |
Temperature Scale | ITS-90 |
Reading Renewal | 4 time/sec |
Operation Temperature and Humidity: | 0°C~50°C (32°F~122°F);<80% RH |
Storage Temperature and Humidity: | -10°C~60°C (-4°F~140°F);<70% RH |
Power Requirement: | SIZE AAA/1.5V/UM-4 x 4PCS |
Battery Life: | Approx. 250 hours/alkaline battery (general pattern) |
Save Power Mode for PTD-12A only) | Approx. 3000 hours/Save electricity pattern one (interval = 15min.) |
Approx. 6 months/Save electricity pattern two (interval = 15min.) | |
Dimension: | 164 mm X 76 mm X 32 mm. |
Weight: (include batteries) | Approx. 415g |